Clinician reliability of eyelid and periorbital measurements in the oculoplastic assessment

Author: Yi Ting Yu
Base Hospital / Institution: Barts and The London Medical School, Queen Mary University of London

ePoster presentation

Abstract ID: 25-429

Purpose

To evaluate inter-clinician reliability amongst senior clinicians in eyelid and periorbital assessments in a busy oculoplastics clinic in a real-world setting. Findings will help enhance the accuracy and reproducibility of eyelid assessments.


Methods

116 patients were prospectively recruited at Moorfields Eye Hospital over a 3-month period. 5 clinicians participated in the study, with each patient assessed by 2 clinicians. Assessed manual oculoplastic measurements included palpebral aperture (PA), margin reflex distances (MRD1, MRD2), upper lid show (ULS), lower scleral show (LSS), skin crease height (SCH), levator function (LF), and canthal distances (IPD, ICD, OCD). Inter-clinician agreement was analysed with Bland-Altman plots.


Results

Bland-Altman analysis demonstrated variable inter-clinician agreement across eyelid measurements. MRD1 (-0.35 to 0.6mm), MRD2 (-0.51 to 0.33mm), LF (-0.36 to 0.08mm) and LSS (-0.09 to 0.15mm) showed strong agreement. In contrast, greater inter-clinician variability was observed for PA (–0.67 to 0.16 mm) and SCH (–1.80 to 1.28 mm), with the largest discrepancies seen in the horizontal facial measurements ICD (–2.18 to 1.15 mm) and OCD (–3.14 to 1.09 mm). Outliers beyond the 95% limits were observed across all parameters, highlighting discrepancies in clinician assessments.


Conclusion

Our findings demonstrate that inter-clinician variability depends on the specific eyelid measurement assessed. MRD1, MRD2, LF, and LSS showed strong inter-rater agreement, whereas greater variability was observed for other parameters, such as the horizontal facial measurements ICD and OCD. Outliers across all parameters highlight that inter-clinician variability is an inherent limitation in current practice for measuring eyelid parameters . These results underscore the need to investigate contributing factors and support the development of more standardised or automated tools to improve measurement consistency.


Additional Authors

First name Last name Base Hospital / Institution
Laura Ah-Kye Moorfields Eye Hospital, London, United Kingdom
Yiu Wai Chan Moorfields Eye Hospital, London, United Kingdom
Nigel Griffith Moorfields Eye Hospital, London, United Kingdom
Katie Myint Moorfields Eye Hospital, London, United Kingdom
Allan Ngheim Moorfields Eye Hospital, London, United Kingdom
Amee Patel Moorfields Eye Hospital, London, United Kingdom
Swan Kang Moorfields Eye Hospital, London, United Kingdom

↑ Back to top